◆ Measuring method: Potassium dichromate high-temperature digestion –colorimetric determination method (improved based on the method described in GB11914-89);
◆ Measuring range: (0-500) mg/l, (0-2000) mg/l &(0-5000) mg/l; these three ranges are available;
◆ Max measuring range: 0-40000 mg/l (after automatic dilution);
◆ Chloride ions interference resistance: Max. 40000mg/l (after automatic dilution);
◆ 检测下线:8mg/l;
◆ Lower limit of detection: 8 mg/l;
◆ Resolution: <1 mg/l;
◆ Accuracy: <10% for standard solutions; 15% for water samples;
◆ Reproducibility: <5%;
◆ Digestion temperature: 165 °C; adjustable;
◆ Digestion time: 15 min; adjustable;
◆ Trouble-free operation hours: ≧720 h;
◆ Span drift: ±5% F.S.;
◆ HMI: 7’’ 70000-color TFT true-color touch screen; resolution: 800*480;
◆ Print: Reserved printer interface to connect an external industrial micro-printer (optional);
◆ Storage: 20000 data; no loss of data in the case of power failure; automatic coverage of earliest data when the memory is full; (the memory with a capacity of 40000 data may be added;)
◆ Communication interface: One-channel RS232 digital interface or RS485; support the MODBUS communication protocol or self-defined protocol; 1-channel analog signal: 4-20 ma (with the range corresponding to 20 ma adjustable);
◆ Pretreatment system: Auto cleaning, back flushing and precision filtration; ensure good representativeness of samples and avoid pipe blockage by large suspended particles, thereby ensuring data continuity (optional);
◆ Overall dimensions: 900*600*450 (mm);
◆ Weight: 50 kg;
◆ Power supply: AC220 V±20%, 50 HZ±1%;
◆ Power: 300 W;
◆ Ambient temperature: 5-40 °C;
◆ Ambient humidity: ≦85%;
◆ Measuring method: Nessler reagent (high-temperature distillation) colorimetric determination method.