◆ Method of measurement according to: GB/T 7493-1987; ◆ Measuring range: 0.00-1.00 mg/L; ◆ Accuracy: ≤±10%; ◆ Reproducibility: ≤±10%; ◆ Measuring cycle: Min. 35 min; ◆ Sampling cycle: Periodic (with the interval adjustable between 10 min and 9999 min) and hourly measuring modes; ◆ Calibration cycle: Adjustable within the range of 1 to 99 days; ◆ Maintenance cycle: Once every month and approx. 30 min each time in general; ◆ Reagent consumption: <RMB0.35 yuan per sample; ◆ Output: One-channel RS-232 &2-channel 4 ma-20 ma; GPRS communication. ◆ Environmental requirements: A temperature-controlled room; recommended temperature: +5 °C-28 °C; Humidity: ≤90% RH (non-condensing); ◆ Power supply: AC220 V±10%, 50 Hz±10%, 5 A; ◆ Size: 1500 mm high x500 mm wide x450 mm deep; ◆ Others: Fault alarm &no loss of data in the case of power failure; Touch screen for display and command input;
In the case of fault reset or power ON after power failure, the analyzer automatically discharges residual reagent inside and restores the working condition;ual reagent inside and restores the working condition;
Measuring principle:
In the phosphoric acid medium with a PH of 1.8±0.3, the nitrite reacts with p-aminobenzenesulfonamide (“sulfanilamide” for short) and produces the diazo salt, and then couples with N-(1-naphthyl)-ethylene diamine and produces a red dye; the maximum absorption is realized at the wavelength of 540 nm.